Abstract & Overview
The PhotoMetrix® application was used as a colorimetric analysis tool to determinate the fluoride and phosphorus ion concentration in treated and natural waters.This research aimed to demonstrate the use of a colorimetric analysis tool using a smartphone application called PhotoMetrix® to determinate the fluoride and phosphorus ion concentration in treated and natural waters.The determinations were made using the conventional spectrophotometric approach [2-(4-sulfophenylazo)-1,8-dihydroxy-3,6-naphthalene disulfonate (SPADNS) reagent for fluoride measurements and the molybdenum blue reaction for phosphorus measurements], and then a smartphone equipped with the PhotoMetrix® application was used to compare the results obtained by the two procedures.An analytical curve was developed, and three triplicate samples were analyzed.The coefficient of determination (R 2 ) for phosphorus and fluoride for the conventional method and for the method using the PhotoMetrix® application were 0.990 and 0.998, and 0.999 and 0.999, respectively.The average results obtained using the application showed satisfactory correspondence values of 99.2% (phosphorus) and 100% (fluoride) compared to the reference method.No significant differences between the two methods of quantification were identified (p>0.05,paired t-test).In conclusion, the results demonstrated the efficiency of this new analytical tool, which allows for quick and accurate analysis of fluoride and phosphorus ion concentrations in the field, as well as on an industrial scale.