Abstract & Overview
The triazine herbicides in seawater were determined by solid phase extraction ultra performance liquid chromatography with electrospray ionization tandem mass spectrometry(UPLC-ESI-MS/MS).UPLC and MS/MS analytical conditions were examined and optimized critically by a series of experiments.Samples were filtrated and purified on the HLB solid phase extraction cartridges.The elution was collected and evaporated to less than1mL at 40℃by nitrogen blow,then added to 1mL acetonitrile and water(1∶1,V/V)and ultrasonic wave extraction for 1min.After filtrated with 0.22μm filter membrane,the residue was separated by AcquityTM UPLC HSS C18column(2.1mm×100mm×1.8μm)using gradient elution separation.The mobile phase was a mixture of A(acetonitrile)and B(5mmol/L ammonium acetate containing 0.1%formic acid)in a flow rate of 0.25mL/min.Finally analytes were confirmed and quantified using MS/MS system in multiple reaction mode with triple quadrapole analyzer using positive polarity mode.The analytes show good linearity in the range of 1ng/L to 50ng/L with correlation coefficient from 0.992to 0.999.The detection limit is 1ng/L,and quantification limit is 2ng/L.The average recoveries at three dose levels(2ng/L,5ng/L and 10ng/L)are 65.8%—103% with relative standard deviations from 4.84%to 15.2%.In addition,the method has merits of simplicity,sensitivity and rapidity,and can be used for simultaneous determination of triazine herbicides in seawater.