Abstract & Overview
The effects of absorption spectrum,the pH of buffer and pyrogallic acid concentration on the autoxidation velocity of pyrogallol system were investigated,and the optimum system of pyorgallol autoxidation spectrophotometry was established according to the reaction rules.The reaction volume of reaction system was 3ml.The optimal determining condition when the wavelength was 319nm were as follows:pH of buffer 7.3,pyrogallic acid concentration 60mmol/L,oxidation velocity 0.060△A/min and reaction linearity time 4min.The optimal determining condition when the wavelength was 420nm were as follows:pH of buffer 7.8,pyrogallic acid concentration 80mmol/L,oxidation velocity 0.050△A/min and reaction linearity time 3min.The new wavelength systems having ultraviolet and visible light could be used for the pyrogallic acid reaction system,which could choose the wavelength according to the sample absorption spectrum.The method could eliminate the effect of internal absorption on the detection limit and the pH was close to pH of actual condition in organism.In addition,the autoxidation velocity was the optimum measure value and had better reaction linearity time.