Dose-area product reduction through a practice implementing low frame rate fluoroscopy and increased collimation during single-vessel percutaneous coronary interventions

Checking your access to this record.

Abstract & Overview

This study aims to investigate the impact of lowering fluoroscopic frame rate and increasing collimation on dose area product (DAP) during single-vessel percutaneous coronary interventions. Two patient groups that underwent such procedures were retrospectively reviewed: one treated with a standard practice employing fluoroscopy at 15 frames/s and usual collimation ( standard group), and the other with a reduced practice employing fluoroscopy at 7.5 frames/s and increased collimation ( reduced group). DAP, patient’s effective dose (ED), fluoroscopy time, and number of cineangiography (cine) frames were compared between groups. Multilinear analysis was performed, with DAP as a primary endpoint, to assess the impact of reduced practice in comparison to that of other variables deemed as key descriptors of radiation, patient, disease, or procedure. Relative to the standard group, the reduced group showed a significant decrease of ~37% in DAP (91.46 ± 45.08 Gy•cm2 vs. 145.23 ± 59.67 Gy•cm2, P < 0.0001) and of ~39% in ED (18.70 ± 9.00 mSv vs. 30.61 ± 12.72 mSv, P < 0.0001). While fluoroscopic time was not statistically different between groups (reduced: 10.3 ± 5.6 min vs. standard: 11.1 ± 5.3 min, P = 0.09), number of cine frames was significantly higher in the reduced (1162 ± 304 vs. 936 ± 363, P < 0.0001) due to interoperator experience variability. Multilinear analysis demonstrated that the adoption of the reduced practice was the most significant contributor to the observed DAP decrease. After adjusting for potential confounders, a reduction of ~39% in DAP was found.

Cite this record

C. Bousis, Thanasis Kosovitsas, Pavlos Karanikis, Anna P. Kotsia, Eleni Tzima, Evgenia Pappa (2024). Dose-area product reduction through a practice implementing low frame rate fluoroscopy and increased collimation during single-vessel percutaneous coronary interventions. Radiation Protection and Environment, 47(2), 121–128. https://doi.org/10.4103/rpe.rpe_13_24

Copy in another style

Bousis, C., Kosovitsas, T., Karanikis, P., Kotsia, A. P., Tzima, E., & Pappa, E. (2024). Dose-area product reduction through a practice implementing low frame rate fluoroscopy and increased collimation during single-vessel percutaneous coronary interventions. Radiation Protection and Environment, 47(2), 121–128. https://doi.org/10.4103/rpe.rpe_13_24
Author names are used as deposited, so check their order before you submit.

Metadata is harvested from open-access indexes and can be incomplete. Check a citation against the published version before relying on it — see our open access & attribution policy.