Abstract & Overview
The differences of the results measured by different principle ins tr ument and how to evaluate the precision are introduced.The particle size distrib utions are determined comparatively by sieving method and laser diffraction meth od.The differences of the results are analysed tentatively.It is showed that the least difference of the particle size distribution results measured by sieving method and laser diffraction method is obtained when the accumulative weight(vol ume)distributions are 50% .The different particle size and shape can lead to th e differences of the results determined by the two methods.