Abstract & Overview
A reversed-phase high-performance liquid chromatography (RP-HPLC) method was developed and validated for the quantification of ferulic acid (FA) in nanostructured lipid carriers (NLCs) and biological matrices, employing a Quality by Design (QbD) framework. Method: Chromatographic separation was achieved on a Phenomenex Luna C18 (2) column (250 × 4.6 mm, 5 µm) using an isocratic mobile phase of methanol and 0.1% formic acid (40:60, v/v) at 1.0 mL/min. The method exhibited excellent linearity (r² > 0.999) over the 2-12 µg/mL range. Sensitivity parameters, including a limit of detection of 0.36 µg/mL and a limit of quantification of 1.12 µg/ mL, demonstrated the method’s robustness. Validation, as per international conference of harmonization (ICH) Q2 (R1 & R2) guidelines, confirmed high precision and accuracy (%RSD < 2%). Stability studies under various stress conditions indicated the method’s suitability for bioanalytical applications. Recovery from NLCs, plasma, and urine ranged from 96-99%. Greenness and blueness metrics confirmed the method’s environmental sustainability, and Six Sigma evaluation affirmed its analytical reliability. This method offers a rapid, sensitive, and eco-friendly approach for FA quantification, supporting its application in pharmaceutical development and neurodegenerative disease research.