We use a few cookies

Essential cookies keep the library working. With your permission, we'd also like to use analytics to see what's useful and improve the experience. Read our Privacy Policy or Cookie Policy.

Skip to main content
Home/e-Resources/Electronics & Telecommunication Engineering/A Multi-Frequency Focused Impedance Measurement System Based on Analogue Synchronous Peak Detection
PeriodicalsElectronics & Telecommunication Engineering

A Multi-Frequency Focused Impedance Measurement System Based on Analogue Synchronous Peak Detection

Checking your access to this record.

Abstract & Overview

Monitoring of anatomical structures and physiological processes by electrical impedance has attracted scientists as it is noninvasive, nonionizing and the instrumentation is relatively simple. Focused Impedance Method (FIM) is attractive in this context, as it has enhanced sensitivity at the central region directly beneath the electrode configuration minimizing contribution from neighboring regions. FIM essentially adds or averages two concentric and orthogonal combinations of conventional Tetrapolar Impedance Measurements (TPIM) and has three versions with 4, 6, and 8 electrodes. This paper describes the design and testing of a multi-frequency FIM (MFFIM) system capable of measuring all three versions of FIM at 8 frequencies in the range 10 kHz—1 MHz. A microcontroller based multi-frequency signal generator and a balanced Howland current source with high output impedance (476 kΩ at 10 kHz and 58.3 kΩ at 1 MHz) were implemented for driving currents into biological tissues with an error <1%. The measurements were carried out at each frequency sequentially. The peak values of the amplified voltage signals were measured using a novel analogue synchronous peak detection technique from which the transfer impedances were obtained. The developed system was tested using TPIM measurements on a passive RC Cole network placed between two RC networks, the latter representing skin-electrode contact impedances. Overall accuracy of the measurement was very good (error <4% at all frequencies except 1 MHz, with error 6%) and the resolution was 0.1 Ω. The designed MFFIM system had a sampling rate of >45 frames per second which was deemed adequate for noninvasive real-time impedance measurements on biological tissues.

Cite this record

Muhammad Abdul Kadir, Adrian J. Wilson, K Siddique-e Rabbani (2021). A Multi-Frequency Focused Impedance Measurement System Based on Analogue Synchronous Peak Detection. Frontiers in Electronics, 2, 791016. https://doi.org/10.3389/felec.2021.791016

Copy in another style

Kadir, M. A., Wilson, A. J., & Rabbani, K. S. (2021). A Multi-Frequency Focused Impedance Measurement System Based on Analogue Synchronous Peak Detection. Frontiers in Electronics, 2, 791016. https://doi.org/10.3389/felec.2021.791016
Author names are used as deposited, so check their order before you submit.

Metadata is harvested from open-access indexes and can be incomplete. Check a citation against the published version before relying on it — see our open access & attribution policy.