We use a few cookies

Essential cookies keep the library working. With your permission, we'd also like to use analytics to see what's useful and improve the experience. Read our Privacy Policy or Cookie Policy.

Skip to main content
Home/e-Resources/Computer Science & Information Technology/Lightweight wrist photoplethysmography for heavy exercise: motion robust heart rate monitoring algorithm
PeriodicalsComputer Science & Information Technology

Lightweight wrist photoplethysmography for heavy exercise: motion robust heart rate monitoring algorithm

Checking your access to this record.

Abstract & Overview

The challenge of heart rate monitoring based on wrist photoplethysmography (PPG) during heavy exercise is addressed. PPG is susceptible to motion artefacts, which have to be mitigated for accurate heart rate estimation. Motion artefacts are particularly apparent for wrist devices, for example, a smart watch, because of the high mobility of the arms. Proposed is a low complexity highly accurate heart rate estimation method for continuous heart rate monitoring using wrist PPG. The proposed method achieved 2.57% mean absolute error in a test data set where subjects ran for a maximum speed of 17 km/h.

Cite this record

Po‐Hsiang Lai, In-Soo Kim (2015). Lightweight wrist photoplethysmography for heavy exercise: motion robust heart rate monitoring algorithm. Healthcare Technology Letters, 2(1), 6–11. https://doi.org/10.1049/htl.2014.0097

Copy in another style

Lai, P., & Kim, I. (2015). Lightweight wrist photoplethysmography for heavy exercise: motion robust heart rate monitoring algorithm. Healthcare Technology Letters, 2(1), 6–11. https://doi.org/10.1049/htl.2014.0097
Author names are used as deposited, so check their order before you submit.

Metadata is harvested from open-access indexes and can be incomplete. Check a citation against the published version before relying on it — see our open access & attribution policy.