We use a few cookies

Essential cookies keep the library working. With your permission, we'd also like to use analytics to see what's useful and improve the experience. Read our Privacy Policy or Cookie Policy.

Skip to main content
Home/e-Resources/Mechanical Engineering/The Development and Performance of an Ambulance Stretcher Suspension
PeriodicalsMechanical Engineering

The Development and Performance of an Ambulance Stretcher Suspension

Checking your access to this record.

Abstract & Overview

The deterioration of the medical condition of stretchered patients during ambulance transport is discussed. The dominant vibration is generally low frequency (below 10 Hz) and in the vertical direction. Assessment of the sensitivity of supine patients to front-to-back vibration is referred to. Methods of alleviating the vibration experienced by stretchered patients are discussed and particular types of stretcher suspension described. This suspension has a constant natural frequency of vertical oscillations over a wide range of patient mass. A stretcher ride height (above the ambulance floor) independent of patient mass is achieved. The stretcher suspension was tested over four types o surface (one ofl-highway), using a commercial vehicle having a chassis similar to that of a commonly used ambulance. Vibration of the vehicle floor and of the stretcher in both the vertical and the pitch senses was measured using accelerometers. Transmissibility of the stretcher suspension, the stretcher frame and the stretcher mattress was obtained. The vertical vibration level (averaged over the four types of surface) experienced by the patient is compared with vibration tolerance levels for supine patients. Vibration inputs are closest in magnitude to the vibration exposure limit in the range 1.5–4 Hz. The transmissibility of the suspension in this range is between −7 and −9 dB. Subjective evaluation of the suspension has been favourable.

Cite this record

David R. Leyshon, Charles W Stammers (1986). The Development and Performance of an Ambulance Stretcher Suspension. Proceedings of the Institution of Mechanical Engineers Part D Transport Engineering, 200(4), 249–257. https://doi.org/10.1243/pime_proc_1986_200_188_02

Copy in another style

Leyshon, D. R., & Stammers, C. W. (1986). The Development and Performance of an Ambulance Stretcher Suspension. Proceedings of the Institution of Mechanical Engineers Part D Transport Engineering, 200(4), 249–257. https://doi.org/10.1243/pime_proc_1986_200_188_02
Author names are used as deposited, so check their order before you submit.

Metadata is harvested from open-access indexes and can be incomplete. Check a citation against the published version before relying on it — see our open access & attribution policy.