Abstract & Overview
Currently, systems based on inertial measurement units (IMUs) are used in order to perform analysis and measurements in the feld of human movement. An IMU has accelerometers, gyros and compasses inside, which allow it to measure and report changes in position and speed associated with weight phenomena experienced by a test mass found in the device reference frame. In order to characterize the system based on IMUs of the company Technaid SL, existing in the Technological Faculty of the Francisco José de Caldas District University, Bogotá, Colombia, it was proposed to generate results of its behavior in a static and dynamic way; for this, data was obtained from the sampling, following an established protocol; the study of the data was then performed by means of statistical tests adequate to the results; fnally, the calculation of absolute error, relative error, repeatability and reproducibility (data that did not exist in the technical characteristics of the equipment and required in motion investigations) was performed. This article gives an account of the procedure carried out in order to arrive at the measurement results: An absolute error of 0,348° in static conditions and 1,53° in dynamic conditions was obtained, which allows us to conclude that the sensors perform with a minimum margin of error both for dynamic and static situations and, therefore, ensure reliablity in its future use.