Abstract & Overview
This paper introduces a method to detect natural gas components by multi-dimensional gas chromatography. The analysis conditions of gas chromatography and valve switching time were confirmed,so it can detect all components with once sample injection. The standard gas was used for qualitative analysis and area normalization method was used for quantitative calculation. This method is highly reproducible and precise, moreover,the refinery gas and thermal simulation products can be detected by this method.