Towards the Reliability Check of Complementary Methods for the Precise Estimation of Peak Efficiency of a Long Run Used Germanium Detector in Low Activity Environmental Measurements

Checking your access to this record.

Abstract & Overview

The gamma-ray spectrometry using long run used hyper pure germanium (HPGe) detector needs a stringent full energy peak (FEP) efficiency calibration for environmental radioactivity measurements. The FEP efficiency calibration was carried out using the complementary methods like analytical estimations and Monte Carlo (MC) simulations. In our work, for the analytical estimations, the relative deviation of the theoretical results with that of experimental FEP efficiency was found to be less than 6%. However, in the case of MC methodologies, an overall estimated standard deviation of 5% could be achievable only with the optimized geometry of the long run used germanium detector for low-activity environmental measurements.

Cite this record

M. Margret, S. Chandrasekaran, Challa Venkata Srinivas, Balasubramanian Venkatraman (2022). Towards the Reliability Check of Complementary Methods for the Precise Estimation of Peak Efficiency of a Long Run Used Germanium Detector in Low Activity Environmental Measurements. Journal of Electrical Electronics Engineering, 1(1). https://doi.org/10.33140/jeee.01.01.13

Copy in another style

Margret, M., Chandrasekaran, S., Srinivas, C. V., & Venkatraman, B. (2022). Towards the Reliability Check of Complementary Methods for the Precise Estimation of Peak Efficiency of a Long Run Used Germanium Detector in Low Activity Environmental Measurements. Journal of Electrical Electronics Engineering, 1(1). https://doi.org/10.33140/jeee.01.01.13
Author names are used as deposited, so check their order before you submit.

Metadata is harvested from open-access indexes and can be incomplete. Check a citation against the published version before relying on it — see our open access & attribution policy.